Descripción

AC/DC current measurements with current ranges up to 2000A AC/DC voltage and resistance measurements DC output 55mm in diameter of measurable conductor Mean value display Sleep function Data hold function Approved for conformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. III 600 V, CAT. II 1000 V)
Description ACA, DCA, ACV, DCV, OHM, Freq. °C Continuity Check
Method of detection Mean Value
Display LCD (Digital display: 3999 counts)
DC Current Range 400/2000A
DC Current Resolution 0.1A
Response Time Approx. 2 seconds
Range Switching Auto-range (on resistance range) Manual-range (on current, voltage range)
Operating temperature and humidity 0 - 40°C, 85% RH or less (no condensation)
Influence of external magnetic field 4A or less at 400 A/m
Influence of conductor position ±(1.5% rdg + 3 dgt) or less
Safety Standard Conforms EN61010-1, EN61010-2-031, EN61010-2-032
Circuit voltage 1000 Vrms or less
Withstanding voltage 5.55 kV AC for one minute
Power Supply LR6 x 2 R6P (SUM-3) x 2
Battery Life Approx. 100 hours (continuous)
Consumed current Approx. 9 mA
Sleep Function Automatically powered down in about 10 minutes after the last switch operation
Diameter of measurable conductor 55 mm at maximum
Dimensions Approx. 105 (W) x 250 (H) x 49 (D) mm
Weight Approx. 530 g
Accessories Batteries, carrying case (93034) Output plug (98012) Test Lead (98011) User's manual (IM CL250)
Peak Hold On all range
Max Hold On current / Voltage range
''